DF flow selector for aerosol therapy

The DF flow selector, used in conjunction with Rs and Qmed® flowmeters, allows to divert the supplied gas through an humidifier or through a hose connector to connect devices used for aerosol therapy.

STRUCTURE
The DF flow selector can be connected through its threaded ring to the flowmeter or supplied directly integrated in the Rs body. The flow can be easily directed to the humidifier for oxygen therapy, through the axial outlet, otherwise it can be diverted through a side hose connector for aerosol therapies.

Tech Info

Sizes (LxWxH)

32x54x69 mm

Weight

0.16 Kg

Supply max. Pressure

500 kPa

Max. Pressure drop
With flow 15 L/min O2

9.20 kPa

Inlet connection

1/4” ISO 3253 F.
M12x1.25 F. • 9/16” UNF EN 13544-2 F.

Outlet connection

1/4” ISO 3253 M.
M12x1.25 M. • 9/16” UNF EN 13544-2 M.

Hose connector for aerosol circuit

Ø 6÷9 mm

Pressure drop at the humidifier side with flow 20 L/min.

17 kPa

Pressure drop at the aerosol side with flow 20 L/min.

34 kPa

Related Products

In compliance with the New Guidelines of the Italian Ministry of Health of 28/03/2013 regarding health advertising concerning medical devices, in vitro medical diagnostic devices and medical surgical facilities, we caution that the information contained in this website is exclusively directed at professional practitioners. All contents are of an informational nature and do not have advertising purposes.